Reproducibility of a Portable Spirometer Based on Differential Pressure Sensor

Kun Zou,Chaomin Ni,Zhijun He,Yongliang Zhang,Zuchang Ma,Jianxia Song,Ke Zhang
DOI: https://doi.org/10.3969/j.issn.1671-7104.2013.01.008
2013-01-01
Abstract:OBJECTIVE:Assessing the reproducibility of a portable spirometer, including reproducibility of inter-observer and day-today.METHODS:Lung ventilation function was performed in 22 healthy volunteers by two observers on the same day and repeated by the first observer after 24h.RESULTS:The inter-observer and day-to-day intra-class correlation coefficients are all higher than 0.75. There are no significant difference between each other. Bland-Altman chart shows good limits of agreement between inter-observer and day-to-day, only scattered data are outside of the limits of agreement.CONCLUSIONS:The portable spirometer shows good inter-observer and day-to-day reproducibility, and can be used for testing lung function in clinical.