Correlation Analysis of Measurements between a Handheld Spirometer and a Conventional Spirometer

Xiang-shao FANG,Zu-yong LI,Guo-jun CHEN
DOI: https://doi.org/10.3969/j.issn.1671-301X.2015.05.008
2015-01-01
Abstract:Objective: To investigate the correlation of measurements between the handheld spirometer and a traditional spirometer. Methods: Measurements obtained from healthy subjects or patients coming to Sun Yat-sen Memorial hospital for pulmonary function tests during October 2014 to February 2015 were analyzed. Results:there was no significantly difference in the accuracy in diagnosing obstructivepulmonary disorder between the two spirometers. The correlation coefficient between the two spirometers for FVC、FEV1、FEV1/FVC、FEF25-75 and PEF was 0.971、0.985、0.966、0.953 and 0.943, respectively(P<0.001 for all). Conclusion: It’s reasonable for a handheld spirometer used for pulmonary function tests in the place of a conventional spirometer regarding to the accuracy and feasibility.