A Correction Method for Ring Artifacts in CT Image Based on Polynomial Fitting

WEI Ying,WU Wen-jin,YANG Min
DOI: https://doi.org/10.3969/j.issn.1674-6864.2010.04.011
2010-01-01
Abstract:Ring artifacts in computed tomography(CT) images are caused by in consistent pixel channel response of detector,As an important factor for quality of CT image,ring arifacts influenc the analysis and the post processing of image.A correction method based on piecewise polynomial fitting is proposed.This method is suited for workpeice which has only one material or when the attenuation coefficient of materials is nearly the same.First,the correction factors of each detector are got with the method of piecewise polynomial fitting,then the nonuniform pixel is correeted;lastly,the CT image is reconstructed.The feasibility of the method is validated by the experiment data.