Transcutaneous Power Link Design for Implanted Systems

Kui Zou,Xiuhan Li,Haixia Zhang
DOI: https://doi.org/10.1109/nems.2009.5068690
2009-01-01
Abstract:Inductive coupling is frequently used for power transfer to implanted electronic devices. Transcutaneous power link design for implanted systems is presented in this paper. MEMS coils which are fabricated with micromanufacturing technology are also tested as receiver coils. The class-E transmitter circuit uses a 5V supply and resonates at a frequency of 1.068MHz. The peak of transmitter coil current is 400mA. Using regular receiver coil with diameter of 3cm and employs high-Q wire can receive 40mW power. Results of using MEMS coil, the maximum receive power is -41dBm.
What problem does this paper attempt to address?