Commissioning of the ERDA-TOF beamline at the ion beam laboratory in Madrid

Andrés Redondo-Cubero,John A. Eliades,Antonio Rodríguez,David García-García,Joonkon Kim
DOI: https://doi.org/10.1016/j.nimb.2024.165334
2024-03-31
Abstract:We report progress for the ERDA time-of-flight beamline at the Centre of Micro-Analysis of Materials in Madrid. This beamline experienced a long shut down and has recently gone through a re-commissioning process. The time stations, based on a twin mirror design, demonstrate a time resolution of 215 ps for 6 MeV He 2+ . Using a standard particle detector for the energy signal we find a mass resolution of 1.4 atomic mass units for 27 Al and 28 Si with 40 MeV Cl 4+ . Detection efficiency needs no corrections for elements heavier than 7 Li. Analysis of a borosilicate reference showed good agreement with the known composition. Based on these results, the facility is considered operational and open for users.
physics, nuclear, atomic, molecular & chemical,nuclear science & technology,instruments & instrumentation