Process development and validation of nextgeneration 3D calibration standards forapplication in optical microscopy

Celina Bozena Hellmich,Lena Heinrich,Matthias Hemmleb,Sebastian Buetefisch,Thomas Weimann,Stefanie Kroker
DOI: https://doi.org/10.1088/1361-6501/ad4dc3
IF: 2.398
2024-05-21
Measurement Science and Technology
Abstract:A scalable wafer-based fabrication process for a new generation of 3D standards enabling the 3D calibration of optical microscopes is presented and validated. The 3D standards are based on step pyramids with several layers in the μm range for height calibration and a system of cylindrical knops distributed across the layers as marks for lateral calibration. This enables calibration for the three coordinate axes and the orthogonality error between them in a single measurement step. The requirements necessary for such a calibration, as optical non-transparency, reproducible flatness of the pyramid step heights and the lowest possible deviations of the lateral marks coordinates, are met by optimising the manufacturing process: The deviation of the height steps distributed over the wafer is ± 3.6 nm and is primarily caused by the layer deposition processes. The lateral manufacturing accuracy was determined using calibrated SEM and show a mean deviation of 20 or 60 nm, depending on the lateral size of the structures. The EBL process and the level of inaccuracy of the SEM standard have an influence on the lateral scaling accuracy. Based on the tactilely generated height values and the coordinates of the mark determined by a calibrated SEM, an example calibration of a CLSM was successfully performed and showed good conformity to conventional calibration techniques.
engineering, multidisciplinary,instruments & instrumentation
What problem does this paper attempt to address?