The Suboptimal QLV Ratio May Indicate the Need for a Left Bundle Branch Area Pacing-Optimized Cardiac Resynchronization Therapy Upgrade

Péter Ezer,Kitti Szűcs,Réka Lukács,Tamás Bisztray,Gábor Vilmányi,István Szokodi,András Komócsi,Attila Kónyi
DOI: https://doi.org/10.3390/jcm13195742
IF: 3.9
2024-09-27
Journal of Clinical Medicine
Abstract:Background: The QLV ratio (QLV/baseline QRS width) is an established intraoperative-measurable parameter during cardiac resynchronization therapy (CRT) device implantation, potentially predicting the efficacy of electrical resynchronization. Methods: Left bundle branch area pacing-optimized CRT (LOT-CRT) is a novel approach with the potential to improve both responder rate and responder level in the CRT candidate patient group, even when an optimal electro-anatomical left ventricular lead position is not achievable. In our observational study, 72 CRT-defibrillator candidate patients with a QRS duration of 160 ± 12 ms were consecutively implanted. Using a QLV-ratio-based implant strategy, 40 patients received a biventricular CRT device (Biv-CRT) with an optimal QLV ratio (≥70%). Twenty-eight patients with a suboptimal QLV ratio ( 80%) in terms of QRS width reduction and LVEF improvement. Conclusions: Our single-center study demonstrated the feasibility of a QLV-ratio-based implantation strategy during CRT implantation. Patients with a LOT-CRT system showed significant improvements, whereas Biv-CRT patients with a super-optimal QLV ratio may not be expected to benefit from an additional LOT-CRT upgrade.
medicine, general & internal
What problem does this paper attempt to address?