Micro-Power Low-Offset Instrumentation Amplifier IC Design for Biomedical System Applications

C.-J. Yen,W.-Y. Chung,M.C. Chi
DOI: https://doi.org/10.1109/tcsi.2004.826208
2004-04-01
Abstract:This work presents a micro-power low-offset CMOS instrumentation amplifier integrated circuit with a large operating range for biomedical system applications. The equivalent input offset voltage is improved using a new circuit technique of offset cancellation that involves a two-phase clocking scheme with a frequency of 20 kHz. Channel charge injection is cancelled by the symmetrical circuit topology. With the wide-swing cascode bias circuit design, this amplifier realizes a very high power-supply rejection ratio (PSRR), and can be operated at single supply voltage in the range between 2.5– 7.5 V. It was fabricated using 0.5-$\mu\hbox {m}$ double-poly double-metal $n$-well CMOS technology, and occupies a die area of 0.2 $\hbox {mm}^{2}$. This amplifier achieves a 160-$\mu\hbox {V}$ typical input offset voltage, 0.05% gain linearity, greater than 102-dB PSRR, an input-referred rms noise voltage of 45 $\mu\hbox {V}$, and a current consumption of 61 $\mu\hbox {A}$ at a low supply voltage of 2.5 V. Experimental results indicate that the proposed amplifier can process the input electrocardiogram signal of a patient monitoring system and other portable biomedical devices.