Fascicle-Selective Ultrasound-Powered Bidirectional Wireless Peripheral Nerve Interface IC
Jianxiong Xu,Jose Sales Filho,Sudip Nag,Liam Long,Eugene Hwang,Camilo Tejeiro,Gerard O'Leary,Yu Huang,Mustafa Kanchwala,Mohammad Abdolrazzaghi,Chenxi Tang,Patty Liu,Yuan Sui,Hao You,Xilin Liu,Jose Zariffa,Roman Genov
DOI: https://doi.org/10.1109/TBCAS.2023.3332258
Abstract:This paper presents an innovative, minimally invasive, battery-free, wireless, peripheral nervous system (PNS) neural interface, which seamlessly integrates a millimeter-scale, fascicle-selective integrated circuit (IC) with extraneural recording and stimulating channels. The system also incorporates a wearable interrogator equipped with integrated machine-learning capabilities. This PNS interface is specifically tailored for adaptive neuromodulation therapy, targeting individuals with paralysis, amputation, or chronic medical conditions. By employing a neural pathway classifier and temporal interference stimulation, the proposed interface achieves precise deep fascicle selectivity for recording and stimulation without the need for nerve penetration or compression. Ultrasonic energy harvesters facilitate wireless power harvesting and data reception, enhancing the usability of the system. Key circuit performance metrics encompass a 2.2 μVrms input-referred noise, 14-bit ENOB, and a 173 dB Schreier figure of merit (FOM) for the neural analog-to-digital converter (ADC). Additionally, the ultra-low-power radio-frequency (RF) transmitter boasts a remarkable 1.38 pJ/bit energy efficiency. In vivo experiments conducted on rat sciatic nerves provide compelling evidence of the interface's ability to selectively stimulate and record neural fascicles. The proposed PNS neural interface offers alternative treatment options for diagnosing and treating neurological disorders, as well as restoring or repairing neural functions, improving the quality of life for patients with neurological and sensory deficits.