The Effectiveness of Ultraviolet Smart D60 in Reducing Contamination of Flexible Fiberoptic Laryngoscopes
Uche C Ezeh,Efstratios Achlatis,Tyler Crosby,Paul E Kwak,Michael S Phillips,Milan R Amin
DOI: https://doi.org/10.1002/lary.30869
Abstract:Objective: To compare the effectiveness of disinfection protocols utilizing a ultraviolet (UV) Smart D60 light system with Impelux™ technology with a standard Cidex ortho-phthalaldehyde (OPA) disinfection protocol for cleaning flexible fiberoptic laryngoscopes (FFLs). Methods: Two hundred FFLs were tested for bacterial contamination after routine use, and another 200 FFLs were tested after disinfection with one of four methods: enzymatic detergent plus Cidex OPA (standard), enzymatic detergent plus UV Smart D60, microfiber cloth plus UV Smart D60, and nonsterile wipe plus UV Smart D60. Pre- and post-disinfection microbial burden levels and positive culture rates were compared using Kruskal-Wallis ANOVA and Fisher's two-sided exact, respectively. Results: After routine use, approximately 56% (112/200) of FFLs were contaminated, with an average contamination level of 9,973.7 ± 70,136.3 CFU/mL. The standard reprocessing method showed no positive cultures. The enzymatic plus UV, microfiber plus UV, and nonsterile wipe plus UV methods yielded contamination rates of 4% (2/50), 6% (3/50), and 12% (6/50), respectively, with no significant differences among the treatment groups (p > 0.05). The pre-disinfection microbial burden levels decreased significantly after each disinfection technique (p < 0.001). The average microbial burden recovered after enzymatic plus UV, microfiber plus UV, and nonsterile wipe plus UV were 0.40 CFU/mL ± 2, 0.60 CFU/mL ± 2.4, and 12.2 CFU/mL ± 69.5, respectively, with no significant difference among the treatment groups (p > 0.05). Micrococcus species (53.8%) were most frequently isolated, and no high-concern organisms were recovered. Conclusion: Disinfection protocols utilizing UV Smart D60 were as effective as the standard chemical disinfection protocol using Cidex OPA. Level of evidence: NA Laryngoscope, 133:3512-3519, 2023.