A new method to measure the mechanical properties of rotary endodontic Ni-Ti files: Measurement of 6-axis force/torque and automatic root canal preparation

Hyungwoo Lee,In-Bog Lee
DOI: https://doi.org/10.4012/dmj.2024-082
2024-10-02
Abstract:The purpose of this study was to develop a new instrument to measure the mechanical properties of rotary endodontic Ni-Ti files (ProTaper Gold F2, ProTaper Ultimate F2, and HyFlex EDM Onefile), and to evaluate the overall utility of the device. The instrument was capable of analyzing the 6-axis force/torque generated by the files during cyclic dynamic movement in a metal curved artificial root canal, and doing automatic cyclic dynamic filing in a resin root canal with a preset vertical force limit by adopting a negative feedback mechanism. By analyzing the 6-axis force/torque, we were able to estimate the position and contact points of the files in the curved root canal. ProTaper Gold showed the highest force/torque in all directions. HyFlex EDM had the highest hysteresis ratio, centering ratio value and NCF (number of cycles to fatigue fracture), while the lowest vertical force.
What problem does this paper attempt to address?