Validity of reducing blood culture incubation time for the BD BACTEC FX blood culture system considering microbiological and clinical aspects

Takehiro Hashimoto,Takaaki Yahiro,Sakirul Khan,Kazunori Kimitsuki,Akira Nishizono,Kazufumi Hiramatsu
DOI: https://doi.org/10.1016/j.jiac.2024.09.011
2024-09-12
Abstract:Introduction: A blood culture (BC) test is vital for diagnosing bacteremia in clinical practice. Although incubation time varies among automated BC systems, 4-5 days is deemed to be sufficient time for the BD BACTEC FX blood culture system. This study compared the clinical and microbiological characteristics of true-positive BC samples on day 5 with those within 4 days to reduce missed true bacteremia cases. Methods: We conducted a retrospective study of blood cultures from hospitalized patients between April 2020 and April 2023 at a tertiary care hospital in Japan. Results: In total, 12,342 BC sets were collected from 6,567 patients. Gram-positive bacilli other than Bacillus spp. and Corynebacterium spp., non-albicans Candida, and yeasts other than Candida spp. were detected more frequently in BC-positive patients on day 5 than in those within 4 days. The gastrointestinal tract was the portal of entry more frequently on day 5 than within 4 days (25 % vs. 4 %, p = 0.006). Conclusion: A 4-day incubation period is sufficient for the BD BACTEC FX blood culture system under routine conditions. However, a 5-day incubation period may be warranted when low pathogenicity is suspected or the gastrointestinal tract is suspected as the portal of entry.