Gas chromatographic assay of underivatized 5,5-diphenylhydantoin (dilantin) in plasma extracts.

D. Sampson,I. Harasymiv,W. Hensley
DOI: https://doi.org/10.1093/CLINCHEM/17.5.382
IF: 12.114
1971-05-01
Clinical Chemistry
Abstract:A simple, rapid, and accurate method is described for determining diphenylhydantoin concentrations in plasma. An internal standard, primidone (primaclone), is added to 2 ml of acidified plasma, and primidone and diphenylhydantoin are extracted into chloroform. The extract, washed with hexane and concentrated, is injected into a gas chromatograph. Extraction and measurement of a single sample requires about 30 min, and as many as 40 samples can be assayed in a working day. Both therapeutic and overdose concentrations of diphenylhydantoin in plasma can be detected and measured. Nine other acidic drugs, added to plasma samples, did not interfere with the analysis.
What problem does this paper attempt to address?