Progress toward absolute calibrations of DC x-ray sources from 1 keV to >30 keV for Z diagnostics applications

Timothy J Webb,Patrick W Lake
DOI: https://doi.org/10.1063/5.0213745
2024-08-01
Abstract:Numerous x-ray diagnostics at the Z facility, time-resolved and time-integrated, would benefit from a local calibrated x-ray source. For many years, Z has used low-power Manson x-ray sources to provide some characterization information for components of Z diagnostics below 30 keV. More recently, we have acquired TruFocus x-ray sources, which operate at voltages up to 100 kV depending on the anode material and model. These DC sources have demonstrated good stability and repeatability. Absolute calibrations of either sources or detectors usually involve another calibrated detector from another source, which has not been readily available locally. Therefore, we have adopted the strategy of characterizing a sensitive semiconductor diode detector by controlling or inferring the active volume. This characterized detector is then used to scale the Manson x-ray spectral shape as measured by an Amptek energy-resolving detector. The result agrees well with the modeled source spectrum. We can apply this technique to the TruFocus as well.