Primary Stability of Dental Implants in Human Jawbones: Experiments & FE Analyses
Patrik Wili,Cedric Rauber,Amal Saade,Salomé Bliggenstorfer,Valentina Ramirez‐Garmendia,Ramon Schweizer,Ainara Irastorza‐Landa,Vivianne Chappuis,Philippe Zysset
DOI: https://doi.org/10.1111/clr.14386
2024-12-15
Clinical Oral Implants Research
Abstract:Objectives Primary stability (PS) is a key factor for promoting osseointegration and long‐term success of dental implants particularly for immediate loading protocols. Beyond the current assessments of PS, an accurate pre‐operative evaluation of PS would contribute to the improvement of surgical planning and treatment outcome. This study used biomechanical testing and homogenized finite element (hFE) analysis to objectively measure PS in the laboratory, and digitally estimate PS from prior μCT reconstructions. Material and Methods Thirty‐five bone samples extracted from the jaws of two donors were examined. Twenty‐two were finally evaluated for PS. After scanning of the samples with μCT, implants were inserted by two experienced surgeons, and various metrics such as μCT‐based bone volume fraction (BV/TV), insertion torque (IT), and resonance frequency analysis (RFA) were assessed to determine PS. Mechanical tests were conducted to measure ultimate force (UFexp) as an objective indicator of PS while the hFE simulations were performed to estimate this same ultimate force (UFsim). Results Higher correlation was found between UFsim and UFexp (R2 = 0.85) than between BV/TV and UFexp (R2 = 0.61), IT and UFexp (R2 = 0.50), and RFA and UFexp (R2 = 0.38). All variables demonstrated a statistically significant linear correlation with UFexp (p
engineering, biomedical,dentistry, oral surgery & medicine