New Poisson denoising method for pulse-count STEM imaging

Taichi Kusumi,Shun Katakami,Ryo Ishikawa,Kazuaki Kawahara,Tiarnan Mullarkey,Julie Marie Bekkevold,Jonathan J P Peters,Lewys Jones,Naoya Shibata,Masato Okada
DOI: https://doi.org/10.1016/j.ultramic.2024.113996
IF: 2.994
Ultramicroscopy
Abstract:With the recent progress in the development of detectors in electron microscopy, it has become possible to directly count the number of electrons per pixel, even with a scintillator-type detector, by incorporating a pulse-counting module. To optimize a denoising method for electron counting imaging, in this study, we propose a Poisson denoising method for atomic-resolution scanning transmission electron microscopy images. Our method is based on the Markov random field model and Bayesian inference, and we can reduce the electron dose by a factor of about 15 times or further below. Moreover, we showed that the method of reconstruction from multiple images without integrating them performs better than that from an integrated image.
What problem does this paper attempt to address?