FPGA-Based Planar Sensor Electrical Impedance Tomography (FPGA-psEIT) System Characterized by Double Feedback Howland Constant- Current Pump and Programmable Front-End Measurement

Isnan Nur Rifai,Prima Asmara Sejati,Shinsuke Akita,Masahiro Takei
DOI: https://doi.org/10.1109/tim.2024.3441023
IF: 5.6
2024-08-25
IEEE Transactions on Instrumentation and Measurement
Abstract:The field-programmable gate array (FPGA)-based planar sensor electrical impedance tomography (FPGA-psEIT) system has been developed for skin sodium content imaging, which is characterized by double feedback Howland constant-current pump (DFHC) and programmable front-end measurement (PFM). In order to ensure the FPGA-psEIT suitability for medical application, DFHC maintains current stability across a broad frequency spectrum, while PFM reduces the measurement error over wide range of impedance level. The hardware performance of FPGA-psEIT system is assessed through the signal-to-noise ratio (SNR) evaluation, which is conducted at various load-impedance levels. Further to validate the system, experiments are undertaken by using calf skin of human subjects to investigate the level of sodium content in the dermis layer. As the result of the performance test of FPGA-psEIT, the SNR is achieved 89 dB at –100 kHz. Lastly, in the human calf experiments, FPGA-psEIT exhibits a consistent tendency to yield higher as compared to the impedance analyzer (IA) with an average difference of 29%. The aforementioned results indicate that FPGA-psEIT system is capable to produce a high-quality image of the conductivity distribution of applied in human calf skin.
engineering, electrical & electronic,instruments & instrumentation
What problem does this paper attempt to address?