A portable instrument for high-speed brain function imaging: FEITER

Hugh McCann,S Talha Ahsan,John L Davidson,Rebecca L Robinson,Paul Wright,Chris J D Pomfrett
DOI: https://doi.org/10.1109/IEMBS.2011.6091777
Abstract:Electrical Impedance Tomography (EIT) can resolve dynamic physiological information deep within human subjects [1], but its sensitivity is challenged in the case of imaging the head [2]. Here, we report a new system called fEITER that has been designed and built to enable functional imaging of the human brain using EIT via scalp-mounted electrodes, integrated with stimulation of evoked responses. Using Field-Programmable Gate Array (FPGA) technology, it provides excellent flexibility in terms of current-pattern excitation and signal processing. The instrument operates at 100 frames/second (fps) with noise of 1 μV on the rms voltage measurements. Clinical trials have been authorized by the UK MHRA and example data from human subjects are presented.