Nonlinear optical properties of human cornea measured by spectral domain Z-scan method

Xi Zeng,Pengfei Qi,Mengdi Li,Zhenzhou Cheng,Lie Lin,Weiwei Liu,Yan Wang,Nan Zhang
DOI: https://doi.org/10.1364/OE.441731
2021-11-08
Abstract:In the myopia correction surgery by femtosecond laser, such as Laser in Situ Keratomileusis (LASIK) and Small Incision Lenticule Extraction (SMILE), the nonlinear refractive index of the cornea may cause the deviation of cutting depth. In order to improve the cornea cutting's accuracy and reduce the possibility of undercorrection, the nonlinear refractive index coefficient n2 of the human cornea must be measured with high accuracy. The spectral domain Z-scan technique can measure n2 of the highly scattering biological tissues with much better signal to noise ratio and thus better accuracy than the conventional methods. In this paper, the n2 coefficient of one ex-vivo human corneal sample was measured by the spectral domain Z-scan technique. Experimental results show that as this corneal sample gradually dehydrates, its n2 coefficients are 1.1 ± 0.1×10-19 m2/W, 1.4 ± 0.2×10-19 m2/W and 1.6 ± 0.2 ×10-19 m2/W respectively for the corneal sample with water contents of 89%, 82%, and 78%. The increase of the water content reduces the value of n2, which is reasonable since the nonlinear refractive index coefficient of water is one order of magnitude smaller than that of the cornea.