Comparison of automated mixer-assisted mini-scale liquid-liquid extraction coupled with full evaporation dynamic headspace extraction with United States Environmental Protection Agency methods for the gas chromatography-mass spectrometric analysis of chlorinated benzenes

Christina Shu Min Liew,Hian Kee Lee
DOI: https://doi.org/10.1016/j.chroma.2021.462131
2021-06-21
Abstract:Three modes of facilitating mini-scale-liquid-liquid-extraction (msLLE) prior to automated integration with full evaporation dynamic headspace (FEDHS) extraction were evaluated in this work. For msLLE, 1.2 mL of dichloromethane (DCM) was added to a conical-bottomed vial containing 7 mL of aqueous sample. The solution was then subjected to three different mixing modes, namely vortex-assistance (where a "whirlpool" was created in the solution), agitation-assistance (where the vial was rotated in circular motion) and quickMix-assistance (where the vial was shaken at a high speed). Vortex-assistance was performed manually while the other two modes were automated using a commercial autosampler. Following this, the DCM extract was transferred automatically to another vial and was then vaporized and sent through a Tenax TA sorbent tube in the FEDHS step. Due to the stronger π interaction between the sorbent and the analytes of interest, the analytes were selectively concentrated while the DCM vapor passed through unhampered. After FEDHS, the analytes were thoroughly desorbed into a gas chromatography-mass spectrometric system for analysis. The applicability of this procedure was validated in the extraction of six chlorinated benzenes (CBs) (1,2-dichlorobenzene, 1,3-dichlorobenzene, 1,4-dichlorobenzene, 1,2,4-trichlorobenzene, 1,2,4,5-tetrachlorobenzne and hexachlorobenzene) from aqueous samples. The quickMix-assisted msLLE-FEDHS approach achieved good absolute extraction recoveries (between 74.2% and 88.7%), low limits of detection (between 0.0006 and 0.0116 µg/L), good linearity (r2≥0.9920), good repeatability (between 1.9% and 8.4%, and good reproducibility (between 9.0% and 13.6%). It was found to be superior to the methods published by the United States Environmental Protection Agency. Five consecutive fully automated quickMix-assisted-msLLE-FEDHS-GC-MS runs spanned only ca. 4 hr.
What problem does this paper attempt to address?