Hemocompatibility of new magnetically-levitated centrifugal pump technology compared to the CentriMag adult pump

David Schibilsky,Setsuo Takatani,Barbara Schibilsky,Tobias Graf,Diana Michels da Silva,Hans Peter Wendel,Meltem Avci-Adali,Christian Schlensak
DOI: https://doi.org/10.1038/s41598-020-78709-0
2020-12-16
Abstract:The specific hemocompatibility properties of mechanical-circulatory-support (MCS)-pump technologies have not previously been described in a comparable manner. We thus investigated the hemocompatibility-indicating marker of a new magnetically-levitated (MagLev) centrifugal pump (MT-Mag) in a human, whole-blood mock-loop for 360 min using the MCS devices as a driving component. We compared those results with the CentriMag adult (C-Mag) device under the same conditions according to ISO10993-4. Blood samples were analyzed via enzyme-linked-immunosorbent-assay (ELISA) for markers of coagulation, complement system, and the inflammatory response. The time-dependent activation of the coagulation system was measured by detecting thrombin-anti-thrombin complexes (TAT). The activation of the complement system was determined by increased SC5b-9 levels in both groups. A significant activation of neutrophils (PMN-elastase) was detected within the C-Mag group, but not in the MT-Mag group. However, the amount of PMN-elastase at 360 min did not differ significantly between groups. The activation of the complement and coagulation system was found to be significantly time-dependent in both devices. However, coagulation activation as determined by the TAT level was lower in the MT-Mag group than in the C-Mag group. This slight disparity could have been achieved by the optimized secondary flow paths and surface coating, which reduces the interaction of the surface with blood.
What problem does this paper attempt to address?