Sequential cyclic-square-wave voltammetric determination of sulfanilamide and ciprofloxacin in environment water samples using a 3D-printed electrochemical device

Marina Di-Oliveira,Diele A. G Araújo,David L.O. Ramos,Lucas V.de Faria,Raquel G. Rocha,Raquel M.F. Sousa,Eduardo M. Richter,Thiago R.L.C. Paixão,Rodrigo A.A. Munoz
DOI: https://doi.org/10.1016/j.electacta.2024.143945
IF: 6.6
2024-02-13
Electrochimica Acta
Abstract:Sulfanilamide (SAD) and ciprofloxacin (CIP) are widely used antibiotics, and their indiscriminate use and disposal contribute to the increase in the contamination levels of aquatic environments. Herein, we propose a 3D-printed carbon black/polylactic acid (CB/PLA) electrochemically activated electrode to determine CIP and SAD concurrently by cyclic square-wave voltammetry (C-SWV) technique. Firstly, square wave voltammetry measurements were performed exploiting the oxidation processes of CIP and SAD (Ox1 CIP and Ox1 SAD at around +1.3 V and +1.1 V vs. Ag|AgCl|KCl (sat.) ). However, the lack of selectivity was verified in the analysis of water samples. Then, the C-SWV strategy was employed and enabled the selective determination of CIP (Ox1 CIP ) in the forward scan and SAD (Red1 SAD ) at around +0.5 V vs. Ag|AgCl|KCl (sat.) ) in the reverse scan. Under selected C-SWV conditions, linear ranges from 4.96 to 40.61 μmol L −1 (r = 0.9977) for CIP (Ox1 CIP ) and 1.99 to 10.78 μmol L −1 (r = 0.9989) for SAD (Red1 SAD ) were obtained. The limits of detection for SAD and CIP were 0.2 and 0.6 μmol L −1 , respectively, and sensitivity was enhanced by 25 % for CIP and 60 % for SAD, employing the C-SWV protocol. In addition, this technique improves the selectivity once the analytes can be determined in distinct potential scans. Satisfactory recovery values ranging from 85 to 112 % were obtained in the analysis of spiked water samples (no dilution step), suggesting no pronounced matrix interference effect, and indicating that the C-SWV methodology ensures good accuracy. This work demonstrates that the union of 3D-printed CB/PLA electrodes and the C-SWV technique can enhance selectivity and sensitivity in the sequential analysis of CIP and SAD, and it is a promising strategy to monitor other pharmacological species in water samples.
electrochemistry
What problem does this paper attempt to address?