Effect of Nebulizer Location and Spontaneous Breathing on Aerosol Delivery During Airway Pressure Release Ventilation in Bench Testing
Hui-Qing Ge,Ji-Mei Wang,Hui-Ling Lin,James B. Fink,Ronghua Luo,Peifeng Xu,Kejing Ying
DOI: https://doi.org/10.1089/jamp.2018.1457
2019-01-01
Journal of Aerosol Medicine and Pulmonary Drug Delivery
Abstract:Background: Airway pressure release ventilation (APRV) maintains a sustained airway pressure over a large proportion of the respiratory cycle, and has a long inspiratory time at high pressure. The purpose of this study was to determine the influence of the APRV with and without spontaneous breathing on albuterol aerosol delivery with a continuous vibrating-mesh nebulizer (VMN) placed at different positions on an adult lung model of invasive mechanical ventilation. Methods: An adult lung model was assembled by connecting a ventilator with a dual-limb circuit to an 8-mm inner diameter endotracheal tube (ETT) and collecting filter attached to a test lung with set compliance of 0.1L/cmH(2)O and resistance of 0.5cmH(2)O/(L<bold>s</bold>). Four ventilator modes were compared: pressure control ventilation (PCV) with no bias flow, PCV with bias flow of 6L/min (PCVBF6), APRV with no spontaneous breaths (APRV), and APRV with spontaneous breath trigger (APRVs). Peak inspiratory pressure, peak end-expiratory pressure, aerosol dose, and nebulization time were similar for all modes. The VMN was placed (1) between Y-piece and inspiratory limb, (2) at the gas outlet of a heated humidifier, and (3) at the gas inlet of a heated humidifier. Albuterol sulfate (5mg/2.5mL) was administered with each run and collected on a filter distal to the ETT. Deposited drug was eluted from each filter (purified water) and analyzed by UV spectrophotometry at 276nm. Analysis of variance [general linear model (GLM) multivariate] was performed using the linear model of multiple variables, significance at p<0.05. Results: Albuterol (in micrograms, meanstandard deviation) delivered was higher with VMN placed at the gas inlet of the humidifier with each mode of ventilation (p<0.01). APRVs has the highest albuterol delivery followed by PCV, PCVBF6, and APRV (1706.2 +/- 60.9g vs. 1490.6 +/- 61.1g vs. 1182.3 +/- 61.4g vs. 1153.1 +/- 99.7g, respectively, p<0.001). The minute volume was positively correlated with the inhaled albuterol dose. Conclusions: Spontaneous breathing increased the albuterol delivery during APRV, compared with APRV alone and PCV modes. Placing the nebulizer proximal to the ventilator was more efficient for all modes tested.