A new method to evaluate fluoroscopic system collimator performance

Joseph R. Steiner,Courtney K. Morrison,Mayur Vaya,Nicholas Bevins,Jeremy Christophel,Matt Vanderhoek
DOI: https://doi.org/10.1002/acm2.14536
2024-10-10
Journal of Applied Clinical Medical Physics
Abstract:Introduction Fluoroscopy uses collimators to limit the radiation field size. Collimators are often evaluated annually during equipment performance evaluations to maintain compliance with regulatory and/or accreditation bodies. A method to evaluate and quantify fluoroscopy collimator performance was developed. Methods A radiation field and displayed image measurement device consisting of radiopaque rulers and radiochromic film strips was placed on the x‐ray source assembly exit window to evaluate fluoroscopy collimator performance. This method was used to evaluate collimator performance on 79 fluoroscopic imaging systems including fixed C‐arms, mobile C‐arms, mini C‐arms, and radiographic fluoroscopic systems. Results The excess length (EL), excess width (EW), and sum EL + EW of the radiation field relative to the displayed image were measured and compared to the limits specified in 21CFR1020.32. Four systems exceeded these limits. Placing the radiation measurement device at the x‐ray source assembly exit window relative to the image receptor cover increased the film exposure rate by a factor up to 14.6. The time required to set up and complete the fluoroscopy collimator performance measurements using this method ranged from 5 to 10 min. Conclusions This method provides an easily implemented quantitative measure of fluoroscopy system collimator performance that satisfies regulatory and accreditation body requirements.
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?