Mirror Furnace for Synchrotron Dark Field X-ray Microscopy Experiments

C. Yildirim,H. Vitoux,L. E. Dresselhaus-Cooper,R. Steinmann,Y. Watier,P. K. Cook,M. Kutsal,C. Detlefs
DOI: https://doi.org/10.48550/arXiv.1912.01255
2019-12-03
Applied Physics
Abstract:We present a multi-purpose mirror furnace designed for synchrotron X-ray experiments. The furnace is optimized specifically for dark-field X-ray microscopy (DFXM) of crystalline materials at the beamline ID06 of the ESRF. The furnace can reach up to ~1600{\deg}C with stability better than 2{\deg}C, and heating and cooling rates up to 30{\deg}C/s. The contact-less design enables samples to be heated either in air or in a controlled atmosphere in a capillary tube. The temperature was calibrated via the thermal expansion of an a-iron grain. Temperature profiles in the y and z axes were measured by scanning a thermocouple through the focal spot of the furnace. In the current configuration of the beamline, the furnace can be used for DFXM, near-field X-ray topography, bright field X-ray nanotomography, high resolution reciprocal space mapping, and limited powder diffraction experiments. As a first application, we present a DFXM case study on isothermal heating of a commercially pure Al single crystal.
What problem does this paper attempt to address?