Hypoglycemic Accuracy and Improved Low Glucose Alerts of the Latest Dexcom G4 Platinum Continuous Glucose Monitoring System

Thomas A Peyser,Katherine Nakamura,David Price,Lucas C Bohnett,Irl B Hirsch,Andrew Balo
DOI: https://doi.org/10.1089/dia.2014.0415
Abstract:Objective: Accuracy of continuous glucose monitoring (CGM) devices in hypoglycemia has been a widely reported shortcoming of this technology. We report the accuracy in hypoglycemia of a new version of the Dexcom (San Diego, CA) G4 Platinum CGM system (software 505) and present results regarding the optimum setting of CGM hypoglycemic alerts. Materials and methods: CGM values were compared with YSI analyzer (YSI Life Sciences, Yellow Springs, OH) measurements every 15 min. We reviewed the accuracy of the CGM system in the hypoglycemic range using standard metrics. We analyzed the time required for the CGM system to detect biochemical hypoglycemia (70 mg/dL) compared with the YSI with alert settings at 70 mg/dL and 80 mg/dL. We also analyzed the time between the YSI value crossing 55 mg/dL, defined as the threshold for cognitive impairment due to hypoglycemia, and when the CGM system alerted for hypoglycemia. Results: The mean absolute difference for a glucose level of less than 70 mg/dL was 6 mg/dL. Ninety-six percent of CGM values were within 20 mg/dL of the YSI values between 40 and 80 mg/dL. When the CGM hypoglycemic alert was set at 80 mg/dL, the device provided an alert for biochemical hypoglycemia within 10 min in 95% of instances and at least a 10-min advance warning before the cognitive impairment threshold in 91% of instances in the study. Conclusions: Use of an 80 mg/dL threshold setting for hypoglycemic alerts on the G4 Platinum (software 505) may provide patients with timely warning of hypoglycemia before the onset of cognitive impairment, enabling them to treat themselves for hypoglycemia with fast-acting carbohydrates and prevent neuroglycopenia associated with very low glucose levels.
What problem does this paper attempt to address?