Comparative Study on Laser Induced Breakdown Spectroscopy Based on Single Pulse and Re-Heating Orthogonal Dual Pulse
Yu Yang,Zhao Nan-jing,Fang Li,Meng De-shuo,Gu Yan-hong,Wang Yuan-yuan,Jia Yao,Ma Ming-jun,Liu Jian-guo,Liu Wen-qing
DOI: https://doi.org/10.3964/j.issn.1000-0593(2017)02-0588-06
2017-01-01
Abstract:In order to improve the detection sensitivity and spectral characteristic of las er-induced breakdown spectroscopy (LIBS),re-heating orthogonal dual-pulse c o nfiguration is adopted to analyze Fe,Pb,Ca and Mg contained in the sample an d soil sample contained different concentrations of heavy mental Cr.Variation relationship between spectral intensity,signal-to-background(SBR) of four c haracteristic spectral lines FeⅠ:404.581 nm,PbⅠ:405.78 nm,CaⅠ:422.67 nm and MgⅠ:518.361 nm and time interval of two laser pulses is discusse d,the best time interval of two laser pulses is obtained 1.0 μs.In the con dition of single pulse and dual-pulse,the enhancement factor of spectral inte nsity of four characteristic spectral lines FeⅠ:404.581 nm,PbⅠ:405.78 nm,CaⅠ:422.67 nm and MgⅠ:518.361 nm is respectively 2.23,2.31,2.42 and 2.10;The time evolution characteristic of spectral intensity of char acteristic spectral lines FeⅠ:404.581 nm and CaⅠ:422.67 nm is considered,and also the variation relationship between spectral acquisition delay time a nd SBR of four characteristic spectral lines,dual-pulse can prolong decay tim e of spectral intensity and improve the SBR of characteristic spectral lines;t ime evolution characteristic of plasma temperature and electron density is compa red in the condition of single pulse and dual-pulse,maximum elevation of plas ma temperature is found to be 730 K,and the maximum increase of electron densi ty is 1.8×1016 cm-3.The limits of detection of heavy mental Cr ar e obtained 38 and 20 μg·g-1 respectively in condition of single and doub le pulse,limit of detection of Cr is reduced approximately 2 times by the cond ition of re-heating orthogonal dual pulse.Results above indicate that re-hea ting orthogonal dual-pulse can improve detection sensibility and spectral chara cteristic of LIBS technique,which provides an effective method for decreasing the limit of detection of elements.