Temperature variations during bone removal procedures similar to third molar extraction using different instruments
GB. Bisson,IM. Sanches,NO. Ciaramicolo,O. Ferreira,O. Ferreira Junior
DOI: https://doi.org/10.1016/j.bjoms.2024.01.011
IF: 2.018
2024-02-03
British Journal of Oral and Maxillofacial Surgery
Abstract:Bone removal is commonly used in the extraction of third molars and the heat generated during the process can interfere with bone repair. The aim of this study was to evaluate the temperature variation presented in bone removal performed with a high speed turbine (400,000rpm), implant motor with straight piece (100,000rpm), low speed micromotor (20,000rpm) and piezoelectric saw (30kHz) in pig mandibles. For this, bone removal was carried out around 20 posterior teeth, under constant saline solution irrigation with syringe and needle. In addition, the time required to perform bone removal was recorded. The results indicated a mean temperature variation and standard deviation of 0,96°C (0,6°C) for the high speed turbine, 1,38°C (0,5°C) with the implant motor, 2,22°C (0,7°C) for the low speed micromotor and 2,90°C (1,3°C) for the piezoeletric. The conventional variance was calculated discounting the variation of time used for bone removal around the teeth. There was a statistically significant difference in temperature variation between the high speed turbine vs micromotor (p=0,00) and the high speed vs piezoeletric (p=0,04). We conclude that there is a statistically significant difference in temperature variation between the instruments used in oral and maxillofacial surgery, with higher rotation speeds resulting in lowest temperature variations and a reduced surgical time.
dentistry, oral surgery & medicine,surgery