Molten salt ion source using glass capillaries as emitter

M. Leger,E. Giglio,S. Guillous,A. Houel
2024-12-12
Abstract:We present a proof of concept of an ion source using a molten salt mixture to produce a beam of desired alkali ions like Cs$^+$ when polarized positively or oxygen complexes like NO$_3^-$ when polarized negatively. Similar to liquid metal ion sources, the ions of the salt are emitted by field evaporation from a Taylor cone formed under the influence of an intense electric field, but unlike liquid metal sources, the meniscus of the Taylor cone is not anchored on the tip of a tungsten needle but rather on the tip of a conical glass capillary containing the molten salt. We succeeded in producing a stable ion beam emission of several $\mu$A for about 50 min. We also discuss the behavior of the molten salt mixture in vacuum conditions, as to our knowledge, molten salts in vacuum conditions are not well documented in the literature, while their properties are essential to the design of the presented ion source.
Materials Science,Instrumentation and Detectors