Ground electron calibration of the Gamma-ray Transient Monitor onboard DRO-A Satellite

Pei-Yi Feng,Zheng-Hua An,Yu-Hui Li,Qi Le,Da-Li Zhang,Xin-Qiao Li,Shao-Lin Xiong,Cong-Zhan Liu,Wei-Bin Liu,Jian-Li Wang,Bing-Lin Deng,He Xu,Hong Lu
2024-11-28
Abstract:The Gamma-Ray Transient Monitor (GTM) is an all-sky monitor onboard the Distant Retrograde Orbit-A (DRO-A) satellite, with the scientific objective of detecting gamma-ray bursts in the energy range of 20 keV to 1 MeV. The GTM is equipped with five Gamma-Ray Transient Probes (GTPs), utilizing silicon photomultiplier (SiPM) arrays coupled with NaI(Tl) scintillators for signal readout. To test the performance of the GTP in detecting electrons, we independently developed a continuous-energy-tunable, low-current, quasi-single-electron accelerator, and used this facility for ground-based electron calibration of the GTP. This paper provides a detailed description of the operational principles of the unique electron accelerator and comprehensively presents the process and results of electron calibration for the GTP. The calibration results indicate that the dead time for normal signals is less than 4 $\mu$s, while for overflow signals, it is approximately 70 $\mu$s, consistent with the design specifications. The GTP's time-recording capability is working correctly, accurately recording overflow events. The GTP responds normally to electrons in the 0.4-1.4 MeV energy range. The ground-based electron calibration validates the design of the GTP and enhances the probe's mass model, laying the foundation for payload development, in-orbit observation strategies, and scientific data analysis.
Instrumentation and Methods for Astrophysics,High Energy Physics - Experiment,Accelerator Physics,Instrumentation and Detectors
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is to verify and calibrate the response performance of Gamma - ray Transient Probes (GTPs) on the Gamma - ray Transient Monitor (GTM) to electrons, in order to ensure their ability to accurately monitor high - energy transient events in deep - space orbits. Specifically, the research focuses on the following aspects: 1. **Development of Electron Calibration Facilities**: In order to test the detection performance of GTPs to electrons, the researchers independently developed a continuous energy - adjustable, low - current, quasi - single - electron accelerator. This accelerator can provide a high - energy electron beam in the ground environment for the calibration of GTPs. 2. **Dead - Time Testing**: The dead - time of GTPs under normal signals and overflow signals was measured experimentally. The results show that for normal signals, the dead - time is less than 4 µs; while for overflow signals, the dead - time is approximately 70 µs, which is consistent with the design specifications. 3. **Energy - Response Testing**: The researchers used an electron accelerator to conduct electron energy - response tests on GTPs in the energy range of 0.4 - 1.4 MeV. The experimental results indicate that GTPs can respond normally to electrons in the energy range of 0.4 - 1.4 MeV and exhibit good energy resolution and detection efficiency in this range. 4. **Comparison between Simulation and Experiment**: Through Geant4 Monte Carlo simulation, the energy deposition of electrons in NaI(Tl) crystals after passing through Be windows and Teflon materials was studied. The simulation results are consistent with the experimental data, verifying the design and performance of GTPs. ### Main Conclusions - **Calibration Results Verify the Design of GTPs**: Through ground - based electron calibration, it is verified that the design of GTPs meets expectations, especially in the performance when encountering complex radiation backgrounds in deep - space environments. - **Establishment of Calibration Database**: Detailed ground - based calibration lays the foundation for the establishment of a calibration database, which is helpful for subsequent in - orbit observation strategies and scientific data analysis. - **Improvement of Probe Quality Model**: The calibration results enhance the quality model of GTPs, ensuring that they can accurately and reliably monitor high - energy transient events in deep - space orbits. ### Formula Display - **Formula for the Bending Radius of Electrons in a Magnetic Field**: \[ \rho[m^{-1}]=\frac{eB_0}{p} = 0.2998\frac{B_0[T]}{p[\text{GeV}/c]} \] where \(\rho\) is the bending radius of electrons, \(B_0\) is the magnetic field intensity, and \(p\) is the momentum of electrons. - **Inverse Landau Function Fitting the Most Probable Energy Peak**: \[ f(x; \text{MPV}, \eta, A)=A\cdot e^{-\frac{1}{2}\left(\frac{\text{MPV}-x}{\eta}+e^{\frac{x - \text{MPV}}{\eta}}\right)} \] where MPV represents the most probable energy deposition in NaI(Tl) crystals, \(\eta\) is the width of the inverse Landau distribution, and \(A\) is the amplitude parameter. - **Quadratic Polynomial Fitting the Relationship between the Most Probable Energy and the Incident Electron Energy**: \[ f(x)=a_0 + a_1\cdot x+a_2\cdot x^2 \] These research results provide solid technical support for GTM to successfully execute scientific missions in deep - space orbits.