Detector performance of the Gamma-ray Transient Monitor onboard DRO-A Satellite

Pei-Yi Feng,Zheng-Hua An,Da-Li Zhang,Chen-Wei Wang,Chao Zheng,Sheng Yang,Shao-Lin Xiong,Jia-Cong Liu,Xin-Qiao Li,Ke Gong,Xiao-Jing Liu,Min Gao,Xiang-Yang Wen,Ya-Qing liu,Xiao-Yun Zhao,Fan Zhang,Xi-Lei Sun,Hong Lu
DOI: https://doi.org/10.1007/s11433-024-2458-9
2024-09-10
Abstract:Gamma-ray Transient Monitor (GTM) is an all-sky monitor onboard the Distant Retrograde Orbit-A (DRO-A) satellite with the scientific objective of detecting gamma-ray transients ranging from 20 keV to 1 MeV. GTM is equipped with 5 Gamma-ray Transient Probe (GTP) detector modules, utilizing the NaI(Tl) scintillator coupled with a SiPM array. To reduce the SiPM noise, GTP makes use of a dedicated dual-channel coincident readout design. In this work, we firstly studied the impact of different coincidence times on detection efficiency and ultimately selected the 500 ns time coincidence window for offline data processing. To test the performance of GTPs and validate the Monte Carlo simulated energy response, we conducted comprehensive ground calibration tests using Hard X-ray Calibration Facility (HXCF) and radioactive sources, including energy response, detection efficiency, spatial response, bias-voltage response, and temperature dependence. We extensively presented the ground calibration results, and validated the design and mass model of GTP detector. These work paved the road for the in-flight observation and science data analysis.
Instrumentation and Methods for Astrophysics,High Energy Physics - Experiment,Nuclear Experiment,Instrumentation and Detectors
What problem does this paper attempt to address?