Detector performance of the Gamma-ray Transient Monitor onboard DRO-A Satellite

Pei-Yi Feng,Zheng-Hua An,Da-Li Zhang,Chen-Wei Wang,Chao Zheng,Sheng Yang,Shao-Lin Xiong,Jia-Cong Liu,Xin-Qiao Li,Ke Gong,Xiao-Jing Liu,Min Gao,Xiang-Yang Wen,Ya-Qing liu,Xiao-Yun Zhao,Fan Zhang,Xi-Lei Sun,Hong Lu
DOI: https://doi.org/10.1007/s11433-024-2458-9
2024-09-10
Abstract:Gamma-ray Transient Monitor (GTM) is an all-sky monitor onboard the Distant Retrograde Orbit-A (DRO-A) satellite with the scientific objective of detecting gamma-ray transients ranging from 20 keV to 1 MeV. GTM is equipped with 5 Gamma-ray Transient Probe (GTP) detector modules, utilizing the NaI(Tl) scintillator coupled with a SiPM array. To reduce the SiPM noise, GTP makes use of a dedicated dual-channel coincident readout design. In this work, we firstly studied the impact of different coincidence times on detection efficiency and ultimately selected the 500 ns time coincidence window for offline data processing. To test the performance of GTPs and validate the Monte Carlo simulated energy response, we conducted comprehensive ground calibration tests using Hard X-ray Calibration Facility (HXCF) and radioactive sources, including energy response, detection efficiency, spatial response, bias-voltage response, and temperature dependence. We extensively presented the ground calibration results, and validated the design and mass model of GTP detector. These work paved the road for the in-flight observation and science data analysis.
Instrumentation and Methods for Astrophysics,High Energy Physics - Experiment,Nuclear Experiment,Instrumentation and Detectors
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is to evaluate and verify the performance of the Gamma - ray Transient Monitor (GTM) carried on the DRO - A satellite. Specifically, the research aims to: 1. **Optimize the time coincidence window of the detector**: By studying the influence of different time coincidence windows on the detection efficiency, a time coincidence window of 0.5 µs was finally selected for offline data analysis. 2. **Calibrate the energy response**: Use the Hard X - ray Calibration Facility (HXCF) and radioactive sources to carry out detailed calibration of the energy response of the Gamma - ray Transient Probe (GTP). This includes the calibration of the energy response in the low - energy range (9 - 160 keV) and the high - energy range (> 160 keV). 3. **Verify the detection efficiency**: Measure and analyze the full - energy - peak detection efficiency of the GTP through ground experiments, and compare it with the simulation results to verify the mass model of the detector. 4. **Test the spatial response, bias voltage response and temperature dependence**: By testing the responses under different positions, bias voltages and temperature conditions, ensure the stability and reliability of the GTP in various environments. ### Main research content 1. **Instrument design and ground calibration**: - **GTP design**: Describes the structure and key components of the GTP, such as the NaI(Tl) crystal and SiPM array. - **Ground calibration setup**: Introduces the method of using the HXCF for hard X - ray calibration and the process of using radioactive sources for high - energy - range calibration. 2. **Ground calibration results**: - **Time coincidence**: Studies the influence of different time coincidence windows on the count rate and detection efficiency, and determines the optimal 0.5 µs time window. - **Energy response**: Shows the fitting results of the energy spectrum, calculates the energy resolution, and analyzes the response characteristics at different energies. - **Detection efficiency**: Indirectly deduces the full - energy - peak detection efficiency of the GTP by comparing with the efficiency of the HPGe detector, and conducts Geant4 simulation verification. - **Spatial response**: Tests the response of the GTP at different positions, and analyzes the influence of crystal luminescence inhomogeneity and light collection variation on the detection performance. ### Conclusion Through comprehensive ground calibration and performance testing, the research team has verified the design and mass model of the GTM, providing a solid foundation for subsequent in - orbit observations and scientific data analysis. These results indicate that the GTM has good detection performance and can effectively monitor gamma - ray transient sources in deep - space orbits. ### Formula summary - **Net energy spectrum calculation formula**: \[ Pha_{\text{net}}=\left(\frac{Pha_X}{T_X - T_{\text{dead},X}}-\frac{Pha_{\text{bg}}}{T_{\text{bg}}-T_{\text{dead},\text{bg}}}\right)\cdot T_X \] - **Energy resolution formula**: \[ \text{Resolution}(E_\gamma)=\frac{2.355\cdot\sigma(E_\gamma)}{E_\gamma} \] \[ Ch(E_\gamma)=q\sqrt{a^2 + b^2 E_\gamma+ c^2 E_\gamma^2} \] - **Influence of X - ray machine on energy resolution**: \[ \sigma_{X-\text{mach}}=\sqrt{\sigma_{\text{beam}}^2-\sigma_{\text{HPGe}}^2} \] \[ \delta_{\text{int}}=2.355\cdot\sqrt{\sigma_{\text{GTP}}^2-\sigma_{X-\text{mach}}^2} \] These formulas help researchers accurately evaluate and optimize the performance of the GTM, ensuring that it can work reliably in the deep - space environment.