Calibration of VELC detectors on-board Aditya-L1 mission

Shalabh Mishra,K. Sasikumar Raja,Sanal Krishnan V U,Venkata Suresh Narra,Bhavana Hegde S,Utkarsha D.,Muthu Priyal V,Pawan Kumar S,Natarajan V,Raghavendra Prasad B,Jagdev Singh,Umesh Kamath P,Kathiravan S,Vishnu T,Suresha,Savarimuthu P,Jalshri H Desai,Rajiv Kumaran,Shiv Sagar,Sumit Kumar,Inderjeet Singh Bamrah,Amit Kumar
DOI: https://doi.org/10.1007/s10686-024-09922-2
2024-03-12
Abstract:Aditya-L1 is the first Indian space mission to explore the Sun and solar atmosphere with seven multi-wavelength payloads, with Visible Emission Line Coronagraph (VELC) being the prime payload. It is an internally occulted coronagraph with four channels to image the Sun at 5000 Å~ in the field of view 1.05 - 3 \rsun, and to pursue spectroscopy at 5303 Å, 7892 Å~ and 10747 Å~ channels in the FOV (1.05 - 1.5 \rsun). In addition, spectropolarimetry is planned at 10747 Å~ channel. Therefore, VELC has three sCMOS detectors and one InGaAs detector. In this article, we aim to describe the technical details and specifications of the detectors achieved by way of thermo-vacuum calibration at the CREST campus of the Indian Institute of Astrophysics, Bangalore, India. Furthermore, we report the estimated conversion gain, full-well capacity, and readout noise at different temperatures. Based on the numbers, it is thus concluded that it is essential to operate the sCMOS detectors and InGaAs detectors at $-5^{\circ}$ and $-17^{\circ}$ C, respectively, at the spacecraft level.
Solar and Stellar Astrophysics,Instrumentation and Methods for Astrophysics
What problem does this paper attempt to address?