Characterization of the BOLDPET optical prototype, an innovative Cherenkov detector for 511 keV $γ$ radiation

Roman Chyzh,Gerard Tauzin,Dominique Yvon,Chi-Hsun Sung,Dominique Breton,Jihane Maalmi,Klaus Schäfers,Christian Weinheimer,Viatcheslav Sharyy
2024-07-02
Abstract:In the present work we describe the design, construction, and testing of the optical prototype developed for the BOLDPET project, with the objective of creating a PET detection module with high spatial and time resolution. The BOLDPET technology uses an innovative detection liquid, trimethylbismuth, for detecting 511 keV $\gamma$-quanta resulting from positron annihilation. The optical signal is exclusively produced through the Cherenkov mechanism, and the produced photons are detected using Planacon microchannel-plate photomultiplier. We achieve an excellent time resolution of 150 ps (FWHM) within a sizable detection volume measuring 55 mm x 55 mm x 25 mm. Through detailed Geant4 simulations, we examine the limiting factors affecting time resolution and explore potential avenues for improvement. Furthermore, we demonstrate the feasibility of coarse 2D localization of interactions using the optical signal alone, achieving a precision of about 5-8 mm (FWHM) within the homogeneous detection volume.
Instrumentation and Detectors,Medical Physics
What problem does this paper attempt to address?
This paper aims to solve the problems of high spatial and temporal resolution in positron emission tomography (PET) technology. Specifically, the paper describes the design, construction, and testing process of the optical prototype developed for the BOLDPET project, whose goal is to create a PET detection module with high spatial and temporal resolution. The BOLDPET technology uses an innovative detection liquid - trimethylbismuth (TMBi) - to detect 511 keV γ - rays produced by positron annihilation. This detection method generates optical signals through the Cherenkov mechanism and uses planar micro - channel plate photomultiplier tubes (MCP - PMT) to detect these photons. ### Main Problems and Goals 1. **Improve Temporal Resolution**: - A temporal resolution of approximately 150 picoseconds (FWHM) was achieved in the study, which was achieved within a relatively large detection volume (55 mm × 55 mm × 25 mm). - Through detailed Geant4 simulations, the factors affecting the temporal resolution were studied, and potential ways for improvement were explored. 2. **Improve Spatial Resolution**: - The paper demonstrates the feasibility of rough 2D positioning using only optical signals, with an accuracy of approximately 5 - 8 millimeters (FWHM). - This method achieves high spatial resolution within the uniform area of the detection volume. ### Technical Details - **Detection Liquid**: Trimethylbismuth (TMBi) is used. This liquid has a short attenuation length (25 mm) and a high photoelectric conversion efficiency (47%), and can convert 511 keV photons into relativistic electrons, thereby generating approximately 20 Cherenkov photons. - **Detector Design**: The detector serves as a time - projection chamber, detecting both optical and charge signals simultaneously. The charge signal is used to accurately determine the 2D position of the γ - interaction, while the optical photons provide accurate time measurements for 3D coordinate reconstruction and time - of - flight (TOF) techniques. - **Readout System**: The micro - channel plate photomultiplier tube (MCP - PMT) Planacon XP85122 from Photonis is used for optical photon detection, and data acquisition and processing are carried out through a specialized readout system. ### Results and Discussion - **Temporal Resolution**: By optimizing the design, especially by reducing the thickness of the TMBi - sensitive volume (from 5 cm to 2.5 cm) and replacing the traditional photomultiplier tube (PMT) with MCP - PMT, the temporal resolution was significantly improved. - **Spatial Resolution**: By analyzing γ - beams at different positions, the ability to perform rough 2D positioning using optical signals was demonstrated, with an accuracy between 5 - 8 millimeters. - **Future Prospects**: Further optimize the time - response function of the PMT to improve the overall time performance. In addition, by measuring the depth - of - interaction coordinate (DOI) and correcting the correlation between the average detection time and DOI, the temporal resolution can be further improved. ### Conclusion The paper demonstrates the high performance of the BOLDPET optical prototype in detecting 511 keV γ - rays, especially in terms of temporal and spatial resolution. Through detailed simulations and experimental verifications, the research team explored the factors affecting the temporal resolution and proposed potential ways for improvement. These results provide an important reference for the development of more accurate brain PET scanners and pre - clinical studies.