Monitoring TES Loop Gain in Frequency Multiplexed Readout

T. de Haan,T. Adkins,M. Hazumi,D. Kaneko,J. Montgomery,G. Smecher,A. Suzuki,Y. Zhou
DOI: https://doi.org/10.1007/s10909-024-03174-2
2024-07-05
Abstract:We present a method for precise monitoring of the loop gain of transition edge sensors (TES) under electrothermal feedback. The measurement is implemented on the ICE DfMux electronics and operates simultaneously with Digital Active Nulling (DAN). It uses one additional bias sinusoid per TES and does not require any additional readout channels. The loop gain monitor is being implemented on the Simons Array and is an integral part of the baseline calibration strategy for the upcoming LiteBIRD satellite.