Study experimental time resolution limits of recent ASICs at Weeroc with different SiPMs and scintillators

Tasneem Saleem,Salleh Ahmad,Jean-Baptiste Cizel,Christophe De La Taille,Maxime Morenas,Vanessa Nadig,Florent Perez,Volkmar Schulz,Stefan Gundacker,Julien Fleury
2023-10-05
Abstract:Medical applications, such as Positron Emission Tomography (PET), and space applications, such as Light Detection and Ranging (LIDAR), are in need of highly specialized ASICs. Weeroc, in collaboration with different partners, is highly involved in developing a new generation of front-end ASICs. In the context of a joined LIDAR project among Weeroc, CNES, and Airbus, Weeroc is working on the development of Liroc, an ASIC for space LIDAR application. Weeroc is also working on advancing ASICs for medical applications with Radioroc under development and intended to be used for PET applications. This study experimentally evaluates the time resolution limits of these ASICs in different configurations, with some of the most recent silicon photomultiplier (SiPM) technologies available on the market, coupled to different scintillation crystals. The best single-photon time resolution (SPTR) was achieved using FBK NUV-HD SiPMs with an FWHM of 90 ps with Liroc and 73 ps with Radioroc. Furthermore, the coincidence time resolution (CTR) of Radioroc was studied with different crystal sizes. Using a large LYSO:Ce,Ca crystal of (3 x 3 x 20 mm3) with Broadcom Near UltraViolet-Metal in Trench (NUV-MT) yields a CTR of 127 ps (FWHM). The best CTR of Radioroc was determined to 83 ps (FWHM) with Broadcom NUV-MT SiPMs coupled to LYSO:Ce,Ca (2 x 2 x 3 mm3) from Taiwan Applied Crystal (TAC).
Instrumentation and Detectors
What problem does this paper attempt to address?