Calibration of a Micromegas-based Gaseous Time Projection Chamber Using Cosmic Ray Muons

Wenming Zhang,Yuanchun Liu,Haiyan Du,Ke Han,Heng Lin,Tao Li,Lingyin Luo,Kaixiang Ni,Yunzhi Peng,Shaobo Wang,Sicheng Wen,Xiyu Yan,Zhiyong Zhang,Wenchang Zhai
DOI: https://doi.org/10.1088/1748-0221/18/07/P07038
2023-07-27
Abstract:We report the calibration of a gaseous Time Projection Chamber based on Micromegas charge readout modules with cosmic ray muons, utilizing their penetrating power and relatively uniform energy deposition per unit length. Muon events were selected through track reconstruction to characterize detector performances, such as the drift velocity, electron lifetime, detector gain, and electric field distortion. The evolution of detector performances with gas purities and electric drift fields over a 50-day data-taking cycle was measured by the muon calibration method. For instance, the drift velocity degraded with gas impurities from ${3.40\pm 0.07 ~ cm/\mu s}$ to ${3.06\pm 0.06 ~ cm/\mu s}$, and then recovered with gas purification. A ${^{137}Cs}$ calibration source was also placed inside the detector as a reference for muon calibrations.
Instrumentation and Detectors,High Energy Physics - Experiment
What problem does this paper attempt to address?