Generating a highly uniform magnetic field inside the magnetically shielded room of the n2EDM experiment
C. Abel,N. J. Ayres,G. Ban,G. Bison,K. Bodek,V. Bondar,T. Bouillaud,D.C. Bowles,G.L. Caratsch,E. Chanel,W. Chen,P.-J. Chiu,C. Crawford,B. Dechenaux,C.B. Doorenbos,S. Emmenegger,L. Ferraris-Bouchez,M. Fertl,P. Flaux,A. Fratangelo,D. Goupillière,W. C. Griffith,D. Höhl,M. Kasprzak,K. Kirch,V. Kletzl,S.V. Komposch,P. A. Koss,J. Krempel,B. Lauss,T. Lefort,A. Lejuez,R. Li,M. Meier,J. Menu,K. Michielsen,P. Mullan,A. Mullins,O. Naviliat-Cuncic,D. Pais,F.M. Piegsa,G. Pignol,G. Quemener,M. Rawlik,D. Rebreyend,I. Rienaecker,D. Ries,S. Roccia,D. Rozpedzik,A. Schnabel,P. Schmidt-Wellenburg,E.P. Segarra,N. Severijns,C.A. Smith,K. Svirina,R. Tavakoli,J. Thorne,S. Touati,J. Vankeirsbilck,R. Virot,J. Voigt,E. Wursten,N. Yazdandoost,J. Zejma,N. Ziehl,G. Zsigmond
2024-10-10
Abstract:We present a coil system designed to generate a highly uniform magnetic field for the n2EDM experiment at the Paul Scherrer Institute. It consists of a main $B_0$ coil and a set of auxiliary coils mounted on a cubic structure with a side length of 273 cm, inside a large magnetically shielded room (MSR). We have assembled this system and characerized its performances with a mapping robot. The apparatus is able to generate a 1 $\mu$ T vertical field with a relative root mean square deviation $\sigma$ ($B_z$)/$B_z$ = 3 $\times$ $10^{-5}$ over the volume of interest, a cylinder of radius 40 cm and height 30 cm. This level of uniformity overcomes the n2EDM requirements, allowing a measurement of the neutron Electric Dipole Moment with a sensitivity better than 1 $\times$ $10^{-27}$ ecm.
Instrumentation and Detectors,Nuclear Experiment