2023-04-14
Abstract:Electrical Impedance Tomography can be cost-effective, portable, non-invasive imaging technique. It has preclinical and a few of them already proven industrial applications. This technique can only recover images of low spatial and contrast resolution, partially due to existing physical models. The capability of discriminating between Impedance profiles in recovery is around 73 percentage. However, similar to other modalities, EITs performance depends on the hardware and recovery algorithm design and operating parameters. This work presents an empirically obtained mutual relation between the hardware design related six independent variables, namely (a) molarity of the coupling media, (b) scanning duration, (c) Size of the object, and (d) parameters defining the size of the scanning assembly (No of electrodes, area of the vessel, and percentage periphery covered by the electrodes), affecting its performance. The expression predicts that the error can be kept under a 10 percentage value in a worst-case scenario if these six parameters are kept under a given range. The root mean square error between the experiment values and predicted values from the presented equation is 1.4053. It is shown that time significantly affects the recovery process compared to other optimizing parameters. It is also shown that the accepted molarity value for the presented system is 2M.
Instrumentation and Detectors,Fluid Dynamics,Medical Physics
What problem does this paper attempt to address?