A novel fast response and radiation-resistant scintillator detector for beam loss monitor

Yuanjing Ji,Zebo Tang,Chen Li,Xin Li,Ming Shao
DOI: https://doi.org/10.1088/1748-0221/12/07/C07042
2023-02-22
Abstract:At high luminosity areas, beam loss monitor with fast response and high radiation resistance is crucial for accelerator operation. In this article, we report the design and test results of a fast response and radiation-resistant scintillator detector as the beam loss monitor for high luminosity colliders, especially at low energy regions such as RFQ. The detector consists of a 2 cm*2 cm 0.5 cm LYSO crystal readout by a 6 mm*6 mm Silicon photomultiplier. Test results from various radioactive sources show that the detector has good sensitivity to photons from tens of keV to several MeV with good linearity and energy resolution (23% for 60 keV {\gamma}-ray). For the field test, two such detectors are installed outside of the vacuum chamber shell of an 800 MeV electron storage ring. The details of the test and results are introduced.