Validation of a Novel Smartphone Pupillometer Application Compared to a Dedicated Clinical Pupillometer in the Measurement of the Pupillary Light Reflex of Healthy Volunteers.

Will Davies,paul middleton,Ilya Buzytsky
DOI: https://doi.org/10.1101/2024.10.18.24315266
2024-10-22
Abstract:Aim: This study aimed to assess the correlation between the PLR curves of 25 healthy volunteers, generated with the MindMirror mobile telephone application, with PLR data from a dedicated clinical pupillometer. Materials & methods: Paired pupillary light reflex curves were recorded from 25 healthy volunteers using the MindMirror mobile telephone application, and a Neuroptics NPi 200 clinical pupillometer. The curves were analysed for correlation using a Pearson correlation coefficient across both curve sets. Results: Close correlation was demonstrated between all parameters except maximum and mean pupillary constriction velocity. These values were consistently lower in the MindMirror derived curves and may be explained by the intensity of the light stimulus available from a mobile telephone being significantly less that that used by the clinical pupillometer. Conclusion: A mobile telephone equipped with a MindMirror mobile phone application may provide a reliable, low cost and widely available alternative to a clinical pupillometer in the assessment of the pupillary light reflex.