Pupil assessment with a new handheld pupillometer in healthy subjects

Ken Asakawa,Mei Imai,Mizuki Ohta,Naomi Kawata,Nanako Kawatsu,Hitoshi Ishikawa
DOI: https://doi.org/10.1007/s10792-022-02387-2
IF: 2.029
2022-07-13
International Ophthalmology
Abstract:To assess the pupil response with a new handheld pupillometer in healthy subjects. Sixty-four eyes of 32 healthy subjects (mean age 21.2 years) were tested. After dark adaptation for 10 min, pupil responses to 1 s red and blue light stimuli at 100 cd/m 2 were measured in the order from right to left eyes with a 1 min interval. The initial pupil size (D1, mm), minimum pupil size (D2, mm), and constriction rate (CR, %) were obtained. Intra-examiner reproducibility was examined using the coefficient of variation (CV, %) and the Bland–Altman plot. Inter-examiner consistency was examined using the interclass correlation coefficient (ICC) and the agreements with a conventional device, by Pearson's correlation coefficient ( r ).