First Commissioning of a Cryogenic Distillation Column for Low Radioactivity Underground Argon

H. O. Back,T. Alexander,A. Alton,C. Condon,E. de Haas,C. Galbiati,A. Goretti,T. Hohmann,An. Ianni,C. Kendziora,B. Loer,D. Montanari,P. Mosteiro,S. Pordes
DOI: https://doi.org/10.48550/arXiv.1204.6061
2012-04-26
Instrumentation and Methods for Astrophysics
Abstract:We report on the performance and commissioning of a cryogenic distillation column for low radioactivity underground argon at Fermi National Accelerator Laboratory. The distillation column is designed to accept a mixture of argon, helium, and nitrogen and return pure argon with a nitrogen contamination less than 10 ppm. In the first commissioning, we were able to run the distillation column in a continuous mode and produce argon that is 99.9% pure. After running in a batch mode, the argon purity was increased to 99.95%, with 500 ppm of nitrogen remaining. The efficiency of collecting the argon from the gas mixture was between 70% and 81%, at an argon production rate of 0.84-0.98 kg/day.