Errata to "A Spherical Coil Array for the Calibration of Whole-Head Magnetoencephalograph Systems"

Yoshiaki Adachi,Daisuke Oyama,Masanori Higuchi,Gen Uehara
DOI: https://doi.org/10.1109/tim.2024.3475788
IF: 5.6
2024-11-20
IEEE Transactions on Instrumentation and Measurement
Abstract:In the above article [1], there is an error in (9). The correct equation is as below: The above error resulted in the calculated values for uncertainty displayed in Fig. 5 and Table III being smaller by 18%. The figure and table reflecting the correct values are as follows:TABLE IIISensor Parameters With Uncertainties channelx(mm)y(mm)z(mm) g(nT/V)CH240.39±0.5820.72±0.56163.84±0.410.1028±0.00840.1757±0.00810.9791±0.00171.176±0.011CH28116.76±0.22-14.51±0.293.61±0.230.9929±0.0008-0.1178±0.00690.0157±0.00691.135±0.008CH55-10.63±0.32116.38±0.2122.02±0.440.0262±0.00780.9993±0.0003-0.0254±0.00031.065±0.008CH115-135.52±0.577.07±0.6278.59±0.60-0.2907±0.00280.3784±0.00960.8788±0.00271.137±0.016CH140-19.74±0.42-116.89±0.3226.20±0.56-0.0926±0.0094-0.9951±0.00110.0352±0.00141.038±0.009The values following the plus-minus signs (±) correspond to uncertainties with a coverage factor . Fig. 5.Fig. 5.