Optimization and Testing of a novel Photoacoustic Spectromicroscopy Cell in the Mid-IR Spectrum

Kirk Michaelian,Chris Kelley,Tor Pedersen,Mark Frogley,Timothy May,Luca Quaroni,Gianfelice Cinque
DOI: https://doi.org/10.48550/arXiv.1706.09844
2018-02-21
Abstract:We have designed, constructed and optimized a novel cell for micro photoacoustic (microPA) spectroscopy (microPAS) as part of a collaboration between the Canadian Light Source and Natural Resources Canada. The cell allows inspection of micrometric samples under a visible light microscope, together with measurement of mid-infrared absorption spectra. Our goal is to introduce a methodology enabling mid-IR investigation of micrometric samples that are inaccessible by other measurement configurations; this includes opaque samples, as well as those that are prone to scattering or possess irregular topography. We have achieved this objective, building and testing a prototype PA spectromicroscopy cell using several different optical sources.
Instrumentation and Detectors
What problem does this paper attempt to address?
The problem that this paper attempts to solve is how to perform mid - infrared spectral measurement on micron - scale samples. Specifically, the author designed, constructed and optimized a new type of micro - photoacoustic spectroscopy microscope (micro PAS) unit, aiming to achieve the measurement of mid - infrared absorption spectra of micron - scale samples. This technique is particularly suitable for samples that are difficult to handle in other measurement configurations, such as opaque samples, easily scattering samples or samples with irregular surfaces. The key contributions of the paper are: 1. **Design and Construction**: A new micro - photoacoustic spectroscopy microscope unit was developed, which is able to observe samples under a visible - light microscope and measure their mid - infrared absorption spectra simultaneously. 2. **Testing and Verification**: The effectiveness and performance of this unit were verified by using multiple optical light sources (including pulsed diode lasers, continuous - wave quantum - cascade lasers, internal FTIR light sources and synchrotron radiation sources). 3. **Application Prospects**: The potential applications of this technique in studying samples that are difficult to be studied by traditional optical mid - infrared microscope configurations, such as highly scattering samples, composite micro - samples, fibers, thick sample slices, etc., were demonstrated. Through these works, the author has proven the feasibility and advantages of the micro - photoacoustic spectroscopy microscope technique in mid - infrared spectral analysis of micron - scale samples, laying the foundation for further research and applications.