A method for characterizing after-pulsing and dark noise of PMTs and SiPMs

A. Butcher,L. Doria,J. Monroe,F. Retiere,B. Smith,J. Walding
DOI: https://doi.org/10.1016/j.nima.2017.08.035
2017-08-22
Abstract:Photo-multiplier tubes (PMTs) and silicon photo-multipliers (SiPMs) are detectors sensitive to single photons that are widely used for the detection of scintillation and Cerenkov light in subatomic physics and medical imaging. This paper presents a method for characterizing two of the main noise sources that PMTs and SiPMs share: dark noise and correlated noise (after-pulsing). The proposed method allows for a model-independent measurement of the after-pulsing timing distribution and dark noise rate.