Development of the Liquid Level Meters for the PandaX Dark Matter Detector

J.Hu,H.Gong,Q. Lin,K. Ni,A. Tan,Y. Wei,M. Xiao,X. Xiao,L. Zhao
DOI: https://doi.org/10.1088/1674-1137/38/5/056002
2013-07-10
Abstract:The two-phase xenon detector is at the frontier of dark matter direct search. This kind of detector uses liquid xenon as the sensitive target and is operated in two-phase (liquid/gas) mode, where the liquid level needs to be monitored and controlled in sub-millimeter precision. In this paper, we present a detailed design and study of two kinds of level meters for the PandaX dark matter detector. The long level meter is used to monitor the overall liquid level while short level meters are used to monitor the inclination of the detector. These level meters are cylindrical capacitors custom-made from two concentric metal tubes. Their capacitance values are read out by a universal transducer interface chip and recorded by the PandaX slow control system. We present the developments that lead to level meters with long-term stability and sub-millimeter precision. Fluctuations (standard deviations) of less than 0.02 mm for the short level meters and less than 0.2 mm for the long level meter were achieved during a few days of test operation.
Instrumentation and Detectors,High Energy Physics - Experiment
What problem does this paper attempt to address?
The problem that this paper attempts to solve is how to achieve accurate monitoring and control of the liquid xenon level in the PandaX (PandaX) dark matter detector. Specifically, the paper focuses on the design and development of two types of liquid level gauges: one long liquid level gauge for monitoring the entire liquid xenon level, and another short liquid level gauge for monitoring the inclination of the liquid - gas interface around the time projection chamber (TPC). These liquid level gauges need to work with sub - millimeter accuracy to ensure the accuracy of the experiment. The paper describes in detail the design, manufacturing methods and performance test results of these liquid level gauges, and proves that they can meet the required stability and accuracy requirements.