Retrieval study of uncemented metal-metal hip prostheses revised for early loosening

Georg Reinisch,Kurt P Judmann,Christian Lhotka,Felix Lintner,Karl A Zweymüller
DOI: https://doi.org/10.1016/s0142-9612(02)00410-6
IF: 14
Biomaterials
Abstract:A tribologic assessment was performed on 22 metal-metal hip prostheses from a single manufacturer, following removal for early aseptic loosening after a mean service life of 32 months (range, 12-59 months). The mean linear wear rate was 7.6 microm/year (range, 2.9-12.8 microm/year). This was below the rates previously observed in other modern metal-metal combinations. A novel contour analysis technique using a coordinate measuring machine showed the mean volumetric wear rate to be 2.02 mm(3)/year (range, 0.55-3.74 mm(3)/year), which corresponds to a mean gravimetric wear rate of 16.9 mg/year (range, 4.6-31.4 mg/year). The mean clearance of 39.8 microm (range, 30-50 microm) was within the optimal range for hard-hard bearing combinations. Evidence of abrasive, adhesive, and third-body wear was found on all bearing surfaces. The tribologic assessment did not indicate manufacturing defects as a cause of early loosening. Equally, third-body wear was too low to be considered a causative factor for early loosening.