FLUKA simulation validation to measure the beam energy of a cyclotron

Jesutofunmi Ayo Fajemisin,Dohyun Kim,David J. Schlyer,Cathy S. Cutler
DOI: https://doi.org/10.1016/j.nimb.2024.165474
2024-07-18
Abstract:FLUKA is a fully integrated particle physics MonteCarlo simulation package which has many applications in high energy experimental physics and engineering. When using a cyclotron for medical radionuclide production, it is important to know that the beam energy is suitable to cause the intended nuclear reaction and minimize unwanted impurities. This work focuses on the comparison of FLUKA as a tool to measure the beam energy of a cyclotron with experimental results using the method of stacked copper foils interspaced by an aluminum degrader. We compared two methods to determine the activity ratio: determining the activity in each foil using the experimentally measured cross-sections from the National Nuclear Data Center (NNDC) cross-section data and calculating the activity in the two Cu foils using the FLUKA program. These results were compared to those reported by Gagnon et al., 2011. The FLUKA results gave energy measurements that were different than the true energy by a percentage in the range of 2.4% to 13%, suggesting that FLUKA is not recommended for pre-calculating the cyclotron beam energy using these low Z materials without experimental confirmation. It also confirms that the experimental cross-section data from the NNDC database is superior for this application giving energy values that ranged from 0.1% to 1.6% from the true energy.
physics, nuclear, atomic, molecular & chemical,nuclear science & technology,instruments & instrumentation
What problem does this paper attempt to address?