Simultaneous determination of organotin, organolead, and organomercury compounds in environmental samples using capillary gas chromatography with atomic emission detection

Yan Liu,V. Lopez-Avila,M. Alcaraz,W. Beckert
DOI: https://doi.org/10.1002/JHRC.1240170705
1994-07-01
Abstract:As part of a continuing evaluation of new analytical and sample preparation techniques conducted by the US Environmental Protection Agency (EPA), the use of capillary gas chromatography with atomic emission detection (GC-AED) for the simultaneous determination of organotin, organolead, and organomercury compounds in environmental samples was investigated. Pentylmagnesium bromide was used to pentylate ionic organotin, organolead, and organomercury compounds; the pentyl derivatives were then separated by GC and determined by AED. Several important GC-AED parameters, including the type of injector inlet, carrier gas flow rate, and helium make-up gas flow rate, were optimized for the simultaneous determination of these organometallic compounds. Their minimum detectable concentrations were approximately 1.0 to 2.5 ng/mL using a 0.5-μL on-column injection. The calibration curves exhibited good linearity between 2.5 and 2500 ng/mL for organotin and organolead compounds, and between 2.5 and 10000 ng/mL for organomercury compounds.
Chemistry
What problem does this paper attempt to address?