Magnetic valves with programmable timing capability for fluid control in paper-based microfluidics

Philip Zwanenburg,Xiao Li,Xinyu Liu
DOI: https://doi.org/10.1109/MEMSYS.2013.6474225
2013-03-07
Abstract:Multi-step analytical tests such as enzyme-linked immunosorbent assay (ELISA) require delivery of multiple fluids into a reaction zone and counting of incubation time at different steps. This paper presents a new type of paper-based magnetic valves that can count the time and turn on or off a fluidic flow accordingly, enabling timed fluid control in paper-based microfluidics. The timing function of these valves is realized using a paper timing channel with an ionic resistor, which can detect the event of a solution flowing through the resistor and trigger an electromagnet (through a simple circuit) to open or close a paper cantilever valve. Based on this principle, we developed normally-open and normally-closed valves with a timing period up to 30.3±2.1 minutes (sufficient for ELISA on paper-based platforms). Using the normally-open valve, we performed an enzyme-based colorimetric reaction commonly used for signal amplification of ELISAs, which requires a timed delivery of enzyme substrate to a reaction zone. This design adds a new fluid-control component to the tool set for developing paper-based microfluidic devices, and has the potential to improve the user-friendliness of these devices.
Materials Science
What problem does this paper attempt to address?